§ 882.1455. Traumatic brain injury eye movement assessment aid.

411 words·~2 min read·/us/cfr/t21/s§ 882.1455·

A research copy — for the controlling text, always check the official state or federal source. Not legal advice.

(a)Identification. A traumatic brain injury eye movement assessment aid is a prescription device that uses a patient's tracked eye movements to provide an interpretation of the functional condition of the patient's brain. This device is an assessment aid that is not intended for standalone detection or diagnostic purposes.
(b)Classification. Class II (special controls). The special controls for this device are:
(1)Clinical performance data under anticipated conditions of use must evaluate tracked eye movement in supporting the indications for use and include the following:
(i)Evaluation of sensitivity, specificity, positive predictive value, and negative predictive value using a reference method of diagnosis;
(ii)Evaluation of device test-retest reliability; and
(iii)A description of the development of the reference method of diagnosis, which may include a normative database, to include the following:
(A)A discussion of how the clinical work-up was completed to establish the reference method of diagnosis, including the establishment of inclusion and exclusion criteria; and
(B)If using a normative database, a description of how the “normal” population was established, and the statistical methods and model assumptions used.
(2)Software verification, validation, and hazard analysis must be performed. Software documentation must include a description of the algorithms used to generate device output.
(3)Performance testing must demonstrate the electrical safety and electromagnetic compatibility
(EMC)of the device.
(4)The patient-contacting components of the device must be demonstrated to be biocompatible.
(5)A light hazard assessment must be performed for all eye-tracking and visual display light sources.
(6)Labeling must include:
(i)A summary of clinical performance testing conducted with the device, including sensitivity, specificity, positive predictive value, negative predictive value, and test-retest reliability;
(ii)A description of any normative database that includes the following:
(A)The clinical definition used to establish a “normal” population and the specific selection criteria;
(B)The format for reporting normal values;
(C)Examples of screen displays and reports generated to provide the user results and normative data;
(D)Statistical methods and model assumptions; and
(E)Any adjustments for age and gender.
(iii)A warning that the device should only be used by trained healthcare professionals;
(iv)A warning that the device does not identify the presence or absence of traumatic brain injury or other clinical diagnoses;
(v)A warning that the device is not a standalone diagnostic; and
(vi)Any instructions to convey to patients regarding the administration of the test and collection of test data. [86 FR 71384, Dec. 16, 2021]
Connections1 cite this
Citation graph
cites case law
§ 882.1455
Traumatic brain injury eye movement assessment aid.
Fed. Reg.×1
Cites 0Cited by 1 across 1 source
★   the supreme law of the land   ★
Don't Tread on Me
E Pluribus Unum — out of many, one

"If you don't know your rights, you don't have any."

Marginalia · a citizen's law index
A research desk, not legal advice. Always read the cited source before relying on a summary.
Questions or an issue? support@self-law.org
disclaimerMarginalia is a research index, not a law firm. Nothing on this site is legal, tax, or financial advice and no attorney–client relationship is formed by using it. Statutes, regulations, and case law change; summaries, search results, AI output, and member posts may be incomplete, out of date, or wrong. Any interpretation drawn from material on this site should be validated by a licensed attorney in your jurisdiction before you act on it.